Advanced Bench top capillary rheometer for research to QC testing
Wide range of shear rates for correlation with real material processing conditions, in one compact system, meeting most test requirements encountered in capillary rheometry. Combining many advanced features associated with larger floor-standing models, the Rosand RH2000 system can be configured for research measurements through to Quality Control applications.
Technical Data
Temperature range
Ambient to 500°C (high temperature option)
5°C to 300°C (low temperature cooling coil option)
Maximum speed
1200mm/min (high speed option)
Maximum force
20kN (option)
![](https://at-messe.netzsch.com/_Resources/Persistent/a/1/6/2/a162df21e37f414ffcdcf8e9427645764f5bbec6/web_Rosand_RH200-1000x1496-600x898.webp)
Number of bores:
Single bore , twin bores
Frame stiffness:
100kN
Dynamic range in speed:
200,000:1 (400,000:1 with high speed option)
Speed uncertainty:
<0.1%
Temperature control range:
<±0.1°C
Bore diameter:
15mm (standard) 9.5mm; 12mm; 19mm; 24mm (bore options)
Barrel bore length:
250mm
Barrel material:
Nitrided steel (standard). Hastelloy; Stainless Steel (barrel options)
Pressure transducer ranges:
30,000psi; 15,000psi; 10,000psi; 5,000psi; 3,000psi; 1,500psi; 1000psi; 500psi; 250psi;
Dies:
Tungsten carbide: precision ±5µm; Hastelloy (optional)
Die diameter:
0.5mm to 2mm (in 0.5mm increments) and 3mm as standard. (Other diameters, including fine bore dies, available on special order)
Atmospheric control:
Nitrogen purge for dry, inert test conditions (option)
![](https://at-messe.netzsch.com/_Resources/Persistent/6/8/b/6/68b600cd72001d17e12d906c69ed0b95328f8ee2/web_Rosand_RH2000_Accessories_02-878x585-600x400.webp)
![](https://at-messe.netzsch.com/_Resources/Persistent/3/8/a/0/38a025b60dcf1154708b53ab1201f3279eb63a16/netzs.ch_contactform.png)
GET IN CONTACT
Scan the QR-Code with your smartphone
You still have questions and would like to be contacted by our sales department? Would you like to receive a quote for the product? Do you need a poster or brochure for one of our products? Just Scan the QR-Code!